F- 2.0.15 [Temperature] Not 35.0 - 39.0 Degrees C And/Or [Conductivity] Not 13.0 - 14.5 Ms; F- 2.0.16 No Flow Errors - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

F- 2.0.15 [TEMPERATURE] NOT 35.0 - 39.0 °C AND / OR [CONDUCTIVITY] NOT 13.0 - 14.5 mS
a) Call the Home screen.
b) Select the [Temperature] window to ENSURE "Temp Setting" is 37.0° C.
c) Press 'CONFIRM'.
d) With the shunt door REMAINING open, call debug screen 0.
e) WITHOUT LOOKING AWAY, watch Flow Error for one
(1) minute. If it EVER = 1, even just once indicates a
Flow Error. Is it EVER = 1?
Yes
Flow Error EVER = 1! Proceed to
No Flow Error ALWAYS = 0! See parts a AND b below:
a) Call debug screen 6.
b) WITHOUT LOOKING AWAY, watch BC Switch (middle column) for five
(5) full minutes OR until if it EVER = 897 or more, even just once,
indicating a Flow problem. TWO (2) possible scenarios:
1)
2) IF BC Switch is NEVER = 897 or more: See procedure number

F- 2.0.16 NO FLOW ERRORS

Call the Home screen. THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) the Temperature window is red: Proceed to
SECTION 4 – TEMPERATURE
2)
IF (and ONLY if) Conductivity is NOT between 13.0 and 14.5 mS: Proceed to
SECTION 5 – CONDUCTIVITY
3)
IF the Temperature window is pale yellow/white AND Conductivity is between 13.0 and 14.5
mS: See (ABOVE) procedure number
IF (and ONLY if) BC Switch is EVER = 897 or more even if only once: Proceed
page 47
to
, procedure number
PROBLEMS.
PROBLEMS.
F- 2.9.2
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 47
, procedure number
F-
5.0.0.
page 214
page 36
(
).
F-
5.0.0.
F- 2.0.16
(page 39).
,
page 317
,
Page 39