F- 9.0.5 GOOD FLOW PUMP PRESSURE / ISOLATE CFS CIRCUIT
a) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')!
b) From the Home screen, set [Dialysate Flow] to 500 ml/min and press 'CONFIRM'.
c) Figure below, insert one of the resistor plugs, from the FOUR-RESISTOR SET, into CFS #10's
distribution board position, "x10, CFS".
d) Call debug screen 10. ACFS should = 0.0?
Yes
ACFS = 0.0! See procedure number F- 9.0.6 (page 102).
No
ACFS DOES NOT = 0.0! See parts a AND b below:
a) ENSURE the resistor plug was placed correctly at position "CFS" before continuing! If
not, repeat procedure number F- 9.0.5 (page 101) from part c.
b) Leaving the resistor plug installed for now, FOUR (4) possible bad components:
1) Actuator-Test Board
4) Distribution board.
1
A) With the machine off, swap in a known good Actuator-Test Board
Dialysis Program ("Select Program" → Dialysis' → 'CONFIRM'); C) If ACFS = 0.0 the
previous Actuator-Test Board is bad.
2
A) With the machine off, swap in a known good Sensor Board
into T and C Mode (refer to OPERATING MODES (page Error! Bookmark not
defined.); C) IMPORTANT! Return to Dialysis Program. If ACFS = 0.0 the previous
Sensor Board is bad.
3
The Sensor Board cable can be checked. NOTE three (3) CFS TRANSDUCER
connections will be checked and proceed to
THE SENSOR BOARD CABLE.
1
OR; 2) Sensor Board
a
To LOCATE the board refer to Figure 4A (page 9)
b
To LOCATE the board refer to Figure 4A (page 9)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
2
OR 3) Sensor Board cable
b
; B) Place the machine
page 524
, SECTION 17 - CHECKING
3
OR;
a
; B) Return to
Page 101