AIR- 1.0.4 AIR SEEN AT L10 / TRACE UPSTREAM FROM CHAMBER #69
Per
Table 10
below, ensuring TMP is between 300 and 360 mmHg, for two (2) minutes, watch for air
starting at Location #11 (L11).
Table 9 – Air Checks (2)
Location
Description
L11
Output Blood Leak Sensor
#8 (see
Figure
L12
Input Blood Leak Sensor #8
(see
L13
Output Valve #25 (see
Figure
75, page 491)
L1
OUTPUT (to machine)
side of Dialysate Filter
#73
(see
If air is seen go to:
L12 (next row)
78, page
494)
L13 (next row)
Figure
78)
L1 (next row)
See (ABOVE) procedure
number
page 492
Figure
74, page
(
491)
Figure 78 – Blood Leak Tubing
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
If NO AIR is seen the leak is from:
Valve #24 / Valve #26 mounting O-Ring(s) (to
LOCATE refer to
DiaSafe
Dialysate Pressure Transducer #9
AIR- 1.0.2
).
Figure
76, page 492) OR
®
Filter Tubing connection
Blood Leak Sensor #8
Valve #25
Page 494