Ve- 5.0.0 Valve Error At Balancing Chamber Cycle 1; Ve- 5.0.1 Valve #32 Unplugged And Valves Do Not Cycle; Ve- 5.0.2 Valve #33 Unplugged And Valves Do Not Cycle - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

VE- 5.0.0 VALVE ERROR AT BALANCING CHAMBER CYCLE 1

Valves #32, #33, #35, and #38 'dots' REMAINING blue. One of these valve
circuits is causing the problem and these procedures isolate between them!
a) At the distribution board (Figure below), unplug
b) Call debug screen 0 and watch the Balancing Chamber Valve 'dots' for
forty-five (45) seconds. Do they begin to cycle between white and blue?
Yes
The Balancing Chamber Valves begin to cycle between white
and blue! Valve #32 OR its blue wire harness is bad.
No Valves DO NOT cycle! See procedure number

VE- 5.0.1 VALVE #32 UNPLUGGED AND VALVES DO NOT CYCLE

a) At the distribution board (Figure above), unplug
b) Watch the Balancing Chamber Valve 'dots' for forty-five (45) seconds. Do they begin to cycle between
white and blue?
Yes
The Balancing Chamber Valves begin to cycle between white and blue! Valve #33 OR its blue
wire harness is bad.
No Valves DO NOT cycle! See procedure number

VE- 5.0.2 VALVE #33 UNPLUGGED AND VALVES DO NOT CYCLE

a) At the distribution board (Figure above), unplug
b) Watch the Balancing Chamber Valve 'dots' for forty-five (45) seconds. Do they begin to cycle between
white and blue?
Valve #32's
Valve #33's
Valve #35's
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
connector.
VE- 5.0.1
(page 201).
connector.
VE- 5.0.2
(page 201).
connector.
Page 201