Fresenius Medical Care 2008K2 Troubleshooting Manual page 616

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Screen on and / or good voltages continued:
IMPORTANT! To prevent damage turn the machine OFF!
A)
B) Your choice, referring to Figure 117 (page 611), plug ONE of the cables back in to where it
belongs as recorded above.
C) Turn the machine on (fan running)!
D) If possible, return to Dialysis Program OR Heat Disinfect.
E) Re-measure +12 and / or +5 volts (within range now?). If the symptom was intermittent call
debug screen 1 to watch 5V Est and / or 12V Est. Allowing sufficient time, does the
symptom you were troubleshooting reoccur?
Yes
Symptom reoccurs! The last device plugged in (or its cable) may be causing the
problem!
No Screen on and/or good voltage(s)! Repeat parts a through e until a cable is plugged
in and the symptom reoccurs indicating the last device plugged in (or its cable) is
causing the problem! If after ALL cables are plugged in AND the problem does not
reoccur either it is very intermittent or reseating a connector solved the problem.
2008K/K2 Troubleshooting Guide
LEFT BLANK INTENTIONALLY
P/N 507298 Rev. H
Page 612