TMP- 3.0.0 ISOLATE DIALYSATE PRESSURE TRANSDUCER

a) Figure below, trace the cable from distribution board position X9, "PDIAL" to the (blue) Dialysate
Pressure Transducer to ENSURE it has not been reversed connected with another hydraulic
component.
b) Call debug screen 1. ENSURE FLWP (upper right column) = 255 constant i.e. Dialysate Flow is "OFF".
c) Remove the red dialyzer connector from the shunt door and place it at dialyzer level.
d) Call debug screen 10. Is ACFS (right corner) between 3.5 and 6.0?
Yes
ACFS between 3.5 and 6.0! See procedure number
No
ACFS is NOT between 3.5 and 6.0! Proceed to
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
TMP- 3.1.0
(page 463).
page 464
, procedure number
TMP-
3.4.0.
Page 462