Co- 4.4.0 Isolate Concentrate Pickup Systems; Co- 4.4.1 Isolate The Connector - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 4.4.0 ISOLATE CONCENTRATE PICKUP SYSTEMS

a) Place the concentrate connectors HARD into their rinse ports!
b) Return to Rinse Program! ALLOW one (1) minute to prime the pumps.
c) Per the Figure (right), place the connector's 'adapter plug'
container of water i.e. do NOT submerse the whole connector!
d) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')!
e) Is AIR FREE concentrate drawn steadily in one direction now?
Yes
Concentrate drawn! There is a problem with the pickup wand
No
See procedure number

CO- 4.4.1 ISOLATE THE CONNECTOR

a) Place the concentrate connectors HARD into their rinse ports
b) Return to Rinse Program!
c) ALLOW one (1) minute to prime the pumps
d) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM').
e) Per the Figure right, remove the connector from the inlet tubing and
submerse the tubing, without the filter, into a container of water.
f)
Is AIR FREE concentrate being drawn steadily in one direction now?
Yes
Concentrate drawn! Per the Figure below, there is a problem with the concentrate connector.
No
See procedure number
CO- 4.4.1
(page 348).
CO- 4.9.0
(page 349).
Figure 59 – Concentrate Connectors
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
ONLY
into a
Page 348