NOTE: These procedures do NOT troubleshoot Vent Tubing overflow!
IMPORTANT! Turn the machine OFF!
A)
B) ENSURE the 'Incoming Water' and 'To Drain' tubing (Figure right)
are not reversed connected!
C) ENSURE the 'To Drain' tubing is not kinked AND, if using a drain
'Quick Connector', it is attached PROPERLY to the station drain!
D) Reattach any 'blown off' tubing and dry the area!
INSTRUCTED OTHERWISE
E) Until
Connector (Figure right) FIGURE104into a bucket!
F) Per the Figure below, was tubing 'blowing off' a UF Check
Valve?
Yes
UF Check Valve tubing 'blowing off'! See procedure
number
LEAKING- 1.0.0
No UF Check Valves okay! See procedure number
LEAKING- 1.1.0
HYDRAULICS TOP VIEW
SECTION 21 – HYDRAULIC LEAKS
, direct the Fluid Sample
(page 554).
page 555
(
).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 553