Co- 1.0.1 Conductivity Is Low/Isolate Bicarb Pump Control; Co- 1.0.2 [Bicarbonate] Window = 0/Acetate Is Selected - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 1.0.1 CONDUCTIVITY IS LOW / ISOLATE BICARB PUMP CONTROL
Call debug screen 0. Figure right, what color is the
WHITE?
1)
IF BLUE: Proceed to
2) IF WHITE: a) Select the 'Dialysate' tab to call the Dialysate screen.
b) Figure below, is the [Bicarbonate] window MORE THAN zero (0)?
CO- 1.0.2 [Bicarbonate] WINDOW = 0 / ACETATE IS SELECTED!
1:34 Acetate is rarely used and uses ONLY one concentrate, ACETATE. Troubleshooting must be
performed using bicarb i.e. the [Bicarbonate] window MUST be more than zero (0). See parts a THROUGH
c below:
known good ACID AND bicarbonate!
a) Firmly connect to
b) Select the screen's 'Conc' button THEN select the
c) Press 'CONFIRM'! ENSURING the [Bicarbonate] window is now MORE THAN zero (0), call debug
screen 0. What color is the
1)
IF BLUE:
ENSURING [Dialysate Flow] is on, allow five (5) minutes. If [Conductivity] remains
less than 13.0 mS see procedure number
2) IF WHITE: Proceed to
page 321
, procedure number
Yes
[Bicarbonate] window more than zero (0)! Proceed to
procedure number
No
[Bicarbonate] window = zero (0)! Acetate concentrate is selected! See
procedure number
Figure 56 – Dialysate Screen
Bic
pump's 'dot' now, BLUE or WHITE?
page 374
, procedure number
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Bic
pump's 'dot', BLUE or
CO-
1.0.3.
CO-
9.0.0.
CO- 1.0.2
(page 320).
ACID
that is attached to the machine.
CO- 1.0.3
(page 321).
CO-
9.0.0.
page 374
,
Page 320