F- 2.6.0 MOTORS OKAY / ISOLATE 'OPTIONAL' DIALYSATE SAMPLER
a) Press the 'Escape' key.
b) At the bottom of the screen, select the Test & Options tab. DO NOT start the tests!
c) Per the Figure below, does the OPTIONAL [Dialysate Sampler] button appear?
Yes
The [Dialysate Sampler] button should NOT appear! Refer to the NOTE below!
No The [Dialysate Sampler] does NOT appear! See procedure number
Test & Options Screen / Optional Dialysate Sampler button
F- 2.7.0 [DIALYSATE SAMPLER] DOES NOT APPEAR
Call the Home screen. Figure right, if the TMP window is RED a TMP alarm
is present. TWO (2) possible scenarios:
1)
IF (and ONLY if) the TMP window is WHITE: See procedure number
2)
IF the TMP window is RED: Perform parts a THROUGH d below:
a) Press and release the 'Reset' key then immediately press and hold it for three (3) seconds.
Allow thirty (30) seconds!
b)
c) If a TMP alarm reoccurs perform parts a and b up to twice more before continuing to part d.
d) Allow twenty (20) seconds then see procedure number
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
F- 2.7.0
The [Dialysate Sampler] button
appears if the Dialysate Sampler
option has been accidentily
turned on. FIGURE3CThis may
cause intermittent flow errors
and/or tubing to blow off. This
option MUST be set to "No" from
Service Mode → Options →
Hardware Options.
F- 2.8.0
(page 34).
F- 2.8.0
(page 34).
(page 33).
Page 33