Nw- 5.0.0 Check Float And Valve #41; Inlet Water Tubing View - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

NW- 5.0.0 CHECK FLOAT AND VALVE #41

IMPORTANT!
a)
TURN THE WATER OFF!
b) Obtain a 1000 ml graduated cylinder.
c) Per the Figure below, locate the clear tubing (from Valve #41) that enters the 'Air Gap Tower' vertically
at the top of the Hydrochamber.
d) Pull the tubing off the 'Air Gap Tower' and direct it into the cylinder.
e) TURN THE WATER ON and measure for one (1) minute!
f)
TWO (2) possible scenarios:
1)
IF (and ONLY if) more than 800 ml collected: See procedure number
2) IF less than 800 ml collected: There is not enough water from the RO OR a there is a restriction
at, to or from Valve #41.
.
Figure 31 – Inlet Water Tubing
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
NW- 5.1.0
(page 154).
Page 153