CO- 10.0.10 ISOLATE TEMP

a) Call debug screen 0. Figure right, ENSURE
Cond
(mS) is within 0.4 of TCD.
b) ENSURE Pre Dialyzer Temp is between 35.5 and 38.5° C.
c) Is Post Dialyzer Temp within 1.2° C of Pre Dialyzer Temp?
Yes
Post Temp within 1.2 of Pre Temp! S
CO- 10.0.11
number
No Post Temp IS NOT within 1.2 of Pre Temp! Proceed to
page 380
CO- 10.0.11 POST TEMP WITHIN 1.2 OF PRE TEMP / ISOLATE POST DIALYZER COND
Post Dialyzer Cond changes but overall is it remaining within +/- 0.4 mS of Pre Dialyzer Cond?
Yes
Post Cond within 0.4 of Pre Cond! See procedure number
No
Post Cond IS NOT within 0.4 of Pre Cond! Proceed to
CO-
10.0.30.
(page 379).
CO-
10.0.20.
, procedure number
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Pre Dialyzer
ee procedure
CO- 10.0.405
page 383
page 385
(
).
, procedure number
Page 379