Uf Check Valves #63 And #64 Illustrations; Section 21 - Hydraulic Leaks - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

NOTE! These procedures do NOT troubleshoot Vent Tubing overflow!
Turn the machine OFF!
A)
B) ENSURE the 'Incoming Water' and 'To Drain' tubing (Figure right)
are not reversed connected!
C) ENSURE the 'To Drain' tubing is not kinked AND, if using a drain
'Quick Connector', it is attached PROPERLY to the station drain!
D) Reattach any 'blown off' tubing and dry the area!
INSTRUCTED OTHERWISE
E) Until
Connector (Figure right) FIGURE104into a bucket!
F) Per
Figure 104
below, was tubing 'blowing off' a UF Check
Valve?
Yes
UF Check Valve tubing 'blowing off'! See procedure
number
LEAKING- 1.0.0
No UF Check Valves okay! Proceed to
number
LEAKING-
HYDRAULICS TOP VIEW
SECTION 21 – HYDRAULIC LEAKS
, direct the Fluid Sample
(page 599).
page 600
1.1.0.
Figure 104 – UF Check Valves #63 AND #64
2008T Troubleshooting Guide
P/N 490292 Rev. A
, procedure
Page 598