Message
Purpose of Message
bibag: PHT Failed
The bibag online Pressure Holding
Test has failed. The machine is in
bypass mode.
bibag: Post Rinse
The machine is rinsing the hydraulics
after emptying the bag.
bibag: Press Calib Err
bibag system pressure calibration
error.
bibag: Press Sensor Err
The bibag connector pressure sensor
is experiencing an error. The machine
is in bypass mode.
169
Troubleshooting
Type
Action Required
Water
If the PHT failed on the "Select Program" screen:
Alarm
If the PHT failed in Dialysis Mode:
Advisory
Advisory only. No action is required.
Advisory
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Water
Open the bibag door, wait 30 seconds to relieve the pressure, and
Alarm
remove the bag. Place the bag back on the connector and close the
door. If the alarm is not cleared, return the blood to the patient if
alarm occurs during treatment. Take the machine out of service and
alert a qualified service technician.
ALARMS
TROUBLE-
SHOOTING
Turn machine power off and back on to rerun the test.
If the failure message is repeated on the next test, take the
machine out of service and alert a qualified service
technician.
Make sure there are no concentrate jugs or other objects
obstructing the bag or pressing against it.
Check the machine for leaks. If no leaks are detected:
Open the bibag door, wait 30 seconds to relieve the
o
pressure, and remove the bag.
Place the bag back on the connector and close the
o
door. The bibag online PHT will run again
automatically.
If the failure message is repeated on the next test:
Discontinue use of the bibag system.
o
Insert the blue bicarbonate connector into an
o
appropriate liquid bicarbonate source in order to
continue the treatment.
Alert a qualified service technician.
o
2008T Machine Operator's Manual P/N 490122 Rev N