Chapter 9: Calibration / adjustment

9.1.12 Part 6: Measuring the volume of the UF pump in liters

ADJ. UF-PUMP VOLUME
UF-Pump 1
back to menu ?
Note:
Check the volume.
If necessary, readjust the UF pump
and repeat the procedure.
9-12
Note
If necessary, change the Single-Needle stroke volume
(see chapter 9.1.10 page 9-10):
– Simultaneously press the  and Start/Stop keys.
– Press the  and  (+ / –) keys to set the stroke volume and confirm
with Start/Stop.
Conf
key
Conf
key
LM
Conf
key
LM
UF I/O
key
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
Esc
pulse-amount = 60
key
Enter the number of strokes
by pressing the +/– keys.
Audio paused
key
ACKNOWLEDGED
After
approx. 3 s
Esc
press uf key
key
Remove the line from the
UF pump, close the T-piece.
Hang the line into a
graduated cylinder.
Swmitch on the UF pump by
pressing the UF I/O key.
uf pulses left = 60
The remaining UF strokes are
indicated on the display.
Display field Ž
indicates the number of
preselected strokes.