Art. Pressure Alarm - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

164
Troubleshooting
Art. Pressure Alarm (with
The pressure inside the arterial drip
the upper Arterial
chamber is above the set alarm limits.
Pressure Alarm limit
flashing)
Art. Pressure Alarm (with
The pressure inside the arterial drip
the lower Arterial
chamber is below the set alarm limits
Pressure Alarm limit
flashing)
2008T Machine Operator's Manual P/N 490122 Rev N
Blood
1) Check arterial and venous tubing for kinked line, clotting or
Alarm
clamps.
2) Ensure that the transducer protector is dry and the monitor line is
open. Replace transducer protector, if necessary.
3) Check for clotted fibers in the dialyzer
4) Check to see if blood flow rate is too high, especially with a post-
pump monitor.
5) Press RESET to reset alarm. If applicable, press RESET again
and hold for two seconds to select new alarm limits.
6) If unable to reset the alarm, return blood to the patient if possible.
Do not return clotted blood to the patient.
7) Take the machine out of service and alert a qualified service
technician.
Blood
1) Check the arterial tubing for kinks, clotting, or clamps.
Alarm
2) Check the needle position and access patency.
3) Ensure that the transducer protector is dry and the monitor line is
open. Replace transducer protector, if necessary.
4) Check to see if blood flow rate is too high, especially with a pre-
pump monitor.
Note: Pre-pump arterial monitoring is very sensitive to
access problems (e.g., access spasms, needle tip
occlusions from patient movement). A slower blood pump
rate will bring the pre-pump arterial pressure up. Assess
whether the patient's access is capable of delivering the
prescribed blood flow.
5) Press the RESET key to reset the alarm. If applicable, press the
RESET key again and hold for two seconds to select new alarm
limits. It may be necessary to start the blood pump at a slower
speed and gradually work up to the prescribed rate. If unable to
reset alarm, return blood to the patient if possible. Do not return
clotted blood to the patient.
6) Take the machine out of service and alert a qualified service
technician.