Chapter 2—Daily Preparation for Treatment
Preparing the Dialysis Delivery System
To prepare the 2008T hemodialysis machine for operation:
1. Press the POWER key on the control panel. The "Select Program" screen (see Figure
20) will appear on the monitor after approximately one minute.
2. Flip down the Touchpad located below the Control Panel Keypad.
Note: If the machine is filled with disinfectant or Rinse is the only option that appears in the
"Select Program" screen, the machine must complete a rinse cycle before being used for
treatment. Select Rinse to start the rinse cycle. Upon completion of rinse cycle test the
machine for any residual disinfectant according to the established guidelines of the facility.
Note: During the power up sequence a message is displayed for a few seconds: "Press
Confirm for Service Mode". If this is done, the machine enters the calibration screens
instead of the "Select Program" screen.
Note: The "Select Program" screen displays any additional software applications (Apps)
that are installed on the 2008T hemodialysis machine (see Figure 20 for an example).
Figure 20 – The Select Program Screen and Touchpad
Note: If the bibag disposable is the bicarbonate source, do not pull the blue bicarbonate
connector from the machine's rinse port. Do the following to prepare the bibag disposable:
1. Remove the white plastic seal from underneath the water and bicarbonate nozzles of
the bibag disposable.
2. Open the bibag door on the machine by lifting up on the dark-gray handle.
3. With the white bibag handle facing outward, hang the bag on the bibag connector
nozzles.
4. Close the door, making sure it latches firmly in place. An audible click means the
door is closed.
Note: The bibag disposable must hang freely below the bibag connector. Make certain that
there are no jugs or other objects obstructing or touching the bibag disposable.
46