Auto Heat Disinfect - Fresenius Medical Care 2008K Operator's Manual

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Auto Heat Disinfect

The Auto Heat Disinfect option is a special feature* that allows the user to program the 2008K
hemodialysis machine to automatically run a heat disinfection program according to a schedule.
In the "Options: Default Settings" screen the user may select a start time (Start Prg) and any
day(s) of the week toggle-button(s).
This program affects only the start time of the Heat Disinfect, all other settings function as usual.
With the machine on and set up for rinse, the Heat Disinfect program will run at the selected
time(s).
Note : If another rinse is running when the Auto Heat Disinfect is set to run, the Auto
Heat Disinfect will start after the first rinse completes.
Setting Auto Heat Disinfect Times
1. Go to Service Mode and select the "Options: Default Settings" screen (see figure 64).
2) Select Auto
Disinfect to
display the "Auto
Disinfect" screen.
*Contact Technical Support to activate the Auto Heat Disinfect feature.
2008K Operator's Manual rev. K
Available Software & Hardware Treatment Options and Default Settings
Figure 64 – Service Mode: Options: Default Settings Screen
183