5. Specifications

Cardinal Health™ NPWT PRO family of devices
Dimensions ............................................................................................................. 6 x 4.3 x 2.75 in. (19.3 x 11.0 x 7.0 cm)
Weight .......................................................................................................................................................0.9 Lbs (0.43Kg)
Therapy Settings .....................................................................................................................50, 75, 100, 125, 150 mmHg
Canister Volume ...............................................................................................................................................300cc/500cc
With respect to electric shock, fire, and mechanical hazards, conforms to IEC60601-1.
IEC Classification
• Medical Equipment
• Equipment not suitable for use in presence of flammable anesthetic mixture with air, oxygen, or nitrous oxide.
• Continuous Operation
• Type B Applied Part
• Class II Internally Powered Equipment
• IPXO
Battery
Duration (Fully Charged) ...............................................................................................................................up to 24 hours
Electrical
External Power Supply Input.....100-240 VAC, 50-60Hz, 200 mA or 12-24 VDC, 850 mA (Optional)
External Power Supply Output ......................................................................................................................5 VDC, 1 Amps
Patient & Enclosure leakage Current .......................................................................................................< 100 Micro amps
Environmental Conditions
Storage Conditions
Temperature Range ................................................................................................................. 10°F (-12°C) to 110°F (43°C)
Relative Humidity Range ..........................................................................................................20 – 95% Non-condensing
Atmospheric Pressure Range ....................................................................................................................50 kPa to 110 kPa
Operating Conditions
Temperature Range .......................................................................................................................40°F (4°C) to 90°F (32°C)
Relative Humidity Range ........................................................................................................... 20 - 75% Non-condensing
Atmospheric Pressure Range ....................................................................................................................50 kPa to 110 kPa
Service life ............................................................................................................................................................... 3 years
CAUTION: Federal law restricts this device to sale by or on the order of a physician.