Chapter 2 - Product Specifications - Cardinal Health Infant Flow SiPAP Operator's Manual

Table of Contents

Advertisement

Operator's Manual

Chapter 2 - Product Specifications

Modes
NCPAP
NCPAP with breath rate monitoring and low rate alarm
BiPhasic (time triggered)
BiPhasic (time triggered) with breath rate monitoring and low rate alarm
BiPhasic tr (patient triggered) with breath rate monitoring, low breath rate
alarm and apnea back up (Comprehensive models only)
Controls
Time High (T-High) – 0.1 – 3.0 seconds
Rate (R)
1-120 (Non-U.S. Configuration Parameters)
1-54 (U.S. Configuration Parameters)
Apnea Interval
(T
apnea
Parameters)
(TLBR) – 10-30 seconds; 5 second intervals (U.S. Configuration
Parameters)
NCPAP / Pres Low flow meter – 0-15 L/min, accuracy ± 15% of selected
output
Pres High flow meter – 0-5 L/min, accuracy ± 15% of selected output
Manual Breath – X 1
%O
– 21 -100%
2
Monitors
CPAP
PEEP
MAP
PIP
%O
2
I:E ratio
Spontaneous rate (Rsp)
Battery charge level
675-101-101
Revision H
) – 10-30 seconds, 5 second intervals (Non-U.S. Configuration
3

Advertisement

Table of Contents
loading

Table of Contents