Operator's Manual
Chapter 2 - Product Specifications
Modes
•
NCPAP
•
NCPAP with breath rate monitoring and low rate alarm
•
BiPhasic (time triggered)
•
BiPhasic (time triggered) with breath rate monitoring and low rate alarm
•
BiPhasic tr (patient triggered) with breath rate monitoring, low breath rate
alarm and apnea back up (Comprehensive models only)
Controls
•
Time High (T-High) – 0.1 – 3.0 seconds
•
Rate (R)
1-120 (Non-U.S. Configuration Parameters)
1-54 (U.S. Configuration Parameters)
•
Apnea Interval
•
(T
apnea
Parameters)
•
(TLBR) – 10-30 seconds; 5 second intervals (U.S. Configuration
Parameters)
•
NCPAP / Pres Low flow meter – 0-15 L/min, accuracy ± 15% of selected
output
•
Pres High flow meter – 0-5 L/min, accuracy ± 15% of selected output
•
Manual Breath – X 1
•
%O
– 21 -100%
2
Monitors
•
CPAP
•
PEEP
•
MAP
•
PIP
•
%O
2
•
I:E ratio
•
Spontaneous rate (Rsp)
•
Battery charge level
675-101-101
Revision H
) – 10-30 seconds, 5 second intervals (Non-U.S. Configuration
3