Infusion Specifications -
Maximum infusion rate can be set as part of the configuration.
0.1ml/h - 150ml/h
0.1ml/h - 300ml/h
0.1ml/h - 600ml/h
0.1ml/h - 900ml/h
0.1ml/h - 1200ml/h
The Volume Infused range is 0.0ml - 9990ml.
Bolus Specifications -
Selected maximum rates are shown below
150ml/h
300ml/h
600ml/h
900ml/h
1200ml/h
The default bolus volume can be set as part of the configuration.
Minimum: 0.1ml;
Maximum 100.0ml
Increments of 0.1ml; default 5.0ml
During BOLUS the pressure limit alarms are temporarily increased to
their maximum level.
Critical Volume -
The bolus which can occur in the event of a single internal fault
condition with a 50 ml syringe is :
Maximum Overinfusion - 0.5ml
Purge Specifications -
The purge rate is limited to the maximum rate for the syringe and can
be set as part of the configuration.
100ml/h - 500ml/h.
The purge volume range is 0.5ml - 5ml.
During PURGE the pressure limit alarms are temporarily increased to
their maximum level.
End Of Syringe Rate -
Stop, KVO (0.1ml/h to 2.5ml/h), or set rate if lower than KVO.
Near End Of Infusion Alarm -
5min to end of infusion, or 10% of syringe volume, whichever is
smaller.
End Of Infusion (EOI) Alarm -
0.5% of syringe volume
Electrical Classification -
Class I product. Continuous Mode Operation, Transportable
Maximum Pumping Pressure Limit -
Highest alarm level 1000mmHg (nominal at L-10)
Occlusion Accuracy (% of full scale)* -
L-0
approx.
50 mmHg 300 mmHg 500 mmHg
Temp.
±18%
23°C
* - Using most common 50ml syringes under normal conditions
(95% confidence / 95% of pumps).
Battery Specifications -
Rechargeable sealed NiMH. Automatically charges when the pump is
connected to AC power.
Battery life is typically 4h from fully charged @ 5.0ml/h & 20ºC under
normal conditions. Charging takes 2½ hours from discharge to 90%
charge.
In TCI mode, a fully charged battery allows at least one full syringe to
be infused.
Memory Retention -
The electronic memory of the pump will be retained for more than 6
months when not powered up.
5ml syringes
10ml syringes
20ml syringes
30ml syringes
50ml syringes
5ml syringes
10ml syringes
20ml syringes
30ml syringes
50ml syringes
Pressure mmHg
L-3
L-5
L-10
approx.
approx.
approx.
1000 mmHg
±21%
±23%
±28%
1000DF00005 Iss. 2

Specifications

System Accuracy (continuous mode ml/h and TIVA) -
Volumetric Mean +/- 2% (nominal).
Derating -
Important: System accuracy is +/-2% typical by volume as measured
using the trumpet curve test method defined in IEC/EN60601-2-24
at rates of 1.0ml/h (23ºC) and above when the pump is used with
the recommended syringes. Differences in factors such as size and
plunger force in compatible syringes can cause variations in accuracy
and trumpet curves. See also trumpet curves section in this manual.
Fuse Type -
2 x T 1.25A, slow blowing.
AC Power Supply -
115 - 230VAC, 50 - 60Hz, 20VA (nominal).
Dimensions -
310 mm (w) x 121 mm (h) x 200 mm (d). Weight: 2.7 kg (excluding
power cable).
Protection against fluid ingress -
IPX1 - Protected against vertically falling drops of water.
Alarm Conditions -
Drive Disengaged
Check Syringe
Near End Of Infusion
AC Power Fail
Attention (Nurse Callback)
Dose Would Exceed
Dose not Permitted
Bolus Dose Under
Concentration not Permitted
Rate not Permitted
Environmental Specifications -
Operating Temperature
Operating Relative Humidity
Operating Atmospheric Pressure
Transport & Storage Temperature
Transport & Storage Relative Humidity
Transport & Storage Atmospheric Pressure
Electrical/Mechanical Safety -
Complies with IEC/EN60601-1 and IEC/EN60601-2-24.
EMC -
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.
Temperature +/- 0.5% (5 - 40ºC)
High Rates +/-2.0% (rates > syringe volume/h eg.
>50ml/h in a 50ml syringe.)
Occlusion
Battery Low / Battery Empty
End of Infusion
Internal Malfunction
Titration not confirmed
Target Would Exceed
Dose Under
Bolus Dose not Permitted
Weight Outside Limit
Bolus Dose Over
29/44
+5°C - +40°C
20% - 90%
700hPa - 1060hPa
-30°C - +50°C
10% - 95%
500hPa - 1060hPa