Status Messages - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

20 M
ICROSTREAM

Status Messages

Message
etCO
<LL
2
etCO
> UL
2
iCO
< LL
2
iCO
> UL
2
RRc < LL
RRc > UL
etCO2 Out of
Range (High)
RRc Out of
Range (High)
etCO
Tubing
2
Disconnected
etCO
Tubing
2
Leak
etCO
Tubing
2
Blocked
etCO
Zeroing
2
etCO
Zero
2
Failed
etCO
Zero
2
Accepted
Calibrate etCO
2
pod
etCO
Calibrating Calibration in progress.
2
etCO
Cal.
2
Accepted
etCO
Cal.
2
Canceled
20-8
®
CO2 M
ET
ONITORING
Condition
Values are outside the current alarm
limits.
The CO
value is out of range (high). Check the patient and treat, if
2
The RRc value is out of range (high).
The FilterLine™ is unplugged.
The sidestream tubing has a leak.
NOTE: Occurs only during power up
or during an auto zero.
The FilterLine™ is obstructed.
Autozero in progress
Autozero unsuccessful.
Autozero completed successfully.
An initial calibration is requested after
the first 1,200 operating hours.
Afterwards, calibration is required
annually or at 4,000 operating hours,
whichever is first.
Calibration process completed
successfully.
Calibration cancelled.
D
/D
ELTA
Suggested Action
• Check the patient and treat, if
necessary.
• Change alarm limits.
• Check accessory and replace, if
necessary.
necessary.
Check sample line and reconnect.
Replacethe tubing.
• Disconnect and reconnect the
FilterLine™.
• Disconnect and replace with new
FilterLine™.
Informational message; no action
required.
• Retry to autozero.
• Return the pod to the hospital's
technical personnel.
Informational message; no action
required
• Recalibrate pod.
• Return the pod to the hospital's
technical personnel.
Informational message; no action
required
XL/K
ELTA
APPA
VF8