Dräger Infinity Delta Instructions For Use Manual page 562

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

B T
D
ECHNICAL
Measurement range --
Beta, Alpha, Theta, Delta Power:
Measurement range --
burst suppression ratio (BSR):
Input signal range:
Bandwidth:
Baseline recovery time:
DC offset:
Sampling rate:
EEG scales:
EEG lead labels:
Trending:
Recordings:
Respiration
Sensing leads:
Measurement method:
Detection threshold:
Measurement range:
Measurement accuracy:
Apnea Detection?
B-36
ATA
Percentage of total power
Range: 0 - 100%
Resolution:1%
Accuracy:±10%
Percentage of time waveform is isoelectric over last 60s
Range: 0 to100 %
Resolution:1 %
Accuracy:±10 %
±1.125 mV
0.5 to 30 Hz
≤5 sec
±200 mV DC
8000/sec
5-500 uV/div
None, Left, Right, Front, Back, Right Front, Right Back, Left
Front, Left Back, Fp, Fp1, Fp2, Fz, F3, F4, F7, F8, Cz, C3,
C4, T7, T8, Pz, P3, P4, P7, P8, O, O1, O2
24 hours
Each EEG channel for the following parameters:
R50 recorder:Analog EEG waveforms
Laser printer:Analog EEG waveforms,
I or II (user-selectable)
Impedance pneumography
0.15Ω to 4.0Ω in manual mode (user adjustment)
Adult and Pediatric: 0.20Ω - 10.5Ω in auto mode
Neonatal:
0 - 155 breaths per min
±1 1/breath/min or 2% of rate (whichever is greater)
Adult and pediatric: No
Neonatal:
D
/D
ELTA
Median frequency, spectral edge frequency,
total Power, beta, alpha, theta, and delta
powers, BSR
Power spectrum
(automatic adjustment)
0.20Ω - 1.5Ω in auto mode
(automatic adjustment)
Yes
XL/K
ELTA
APPA
VF8