Overview; Precautions - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

20 M
ICROSTREAM

Overview

®
The Infinity
the monitor to measure and display sidestream etCO
CO
) and RRc (Respiration Rate). The Microstream pod cannot be used
2
simultaneously with either an etCO
The Microstream pod is intended for use with adult, pediatric, and neonatal patients
with appropriate accessories (see page C-24). The pod uses a fixed sample flow rate of
50 mL/min. and requires a proprietary sample line (accessory) and connects to the
monitor via PodCom.

Precautions

Refer to the general precautions section for etCO
CAUTION:
NOTE:
20-2
®
CO2 M
ET
etCO
Microstream
2
To avoid blocking the sampling tube, do not attempt to sterilize the
Microstream pod by immersing it in liquids.
Do not obstruct the exhaust port on the Microstream pod because it prevents
air flow.
The pod and associated accessories that have patient contact are free of latex.
The pod automatically compensates for atmospheric pressure within the defined
operating ranges listed in appendix B.
The Infinity etCO
Microstream SmartPod meets its accuracy specifications
2
within 1.1 minutes after the monitor is powered on.
D
/D
ELTA
ONITORING
®
®
SmartPod
, when connected to a monitor, allows
(end-tidal CO
2
pod or an etCO
/respiratory mechanics pod.
2
2
on page 19-2.
2
XL/K
ELTA
APPA
), iCO
(inspired
2
2
VF8