Dräger Infinity Delta Instructions For Use Manual page 573

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Alarm limit resolution:
Apnea detection?
RRc range (pod):
Rise time:
Delay time:
Total system response time: Rise time plus delay time
Cross-sensitivity Compensation Error: Gas measurement accuracy was tested at the gas
levels listed in the following table. The additional worst case error when compensation is correctly
selected (for barometric pressure, O2, and N2O) for the actual fractional gas constituents that are
present is:
0 to 40 mmHg: ± 2 mmHg additional error
41 to 70 mmHg: ± 3 % additional error
71 to 99 mmHg: ± 5 % additional error
Additional notes regarding cross-sensitivity compensation errors:
Xenon and Helium: Constantly bias etCO2 values >= 38 mmHg by up to –5 mmHg
Desflurane: At concentrations > 5%, constantly biases etCO2 values >= 38 mmHg by up to
+4 mmHg
Gas or Vapor
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Metered dose inhaler
propellants
Desflurane
Ethanol
Isopropanol
VF8
D
etCO
, iCO
: 0.1 mmHg (0.01 kPa)
2
2
RRc: 1 breath/min
Module:
Adult and pediatric:No
Neonatal:Yes
Pod: Yes in all three patient categories
Mainstream:0-149 breaths/min
Sidestream:0-69 breaths/min
Accuracy:±1 breath/min
Mainstream:<100 ms
Sidestream:<200 ms
Mainstream:<100 ms
Sidestream:<450 ms
Quantitative Effects Measured At
Gas Level (Gas levels in % volume fraction)
60
4
5
5
5
80
50
Not specified for use
15
0.1
0.1
/D
XL/K
ELTA
ELTA
M
ONITORING
APPA
S
PECIFICATIONS
B-47