29 C
LEANING AND
To clean the MPod – Quad Hemo
1. Disconnect the MPod – Quad Hemo from the bedside monitor.
2. Clean the MPod – Quad Hemo with a gauze pad moistened with soapy water
or with an approved cleaning agent.
3. Dry thoroughly with a lint-free cloth.
To disinfect the MPod – Quad Hemo
1. Disconnect the MPod – Quad Hemo from the bedside monitor.
2. Disinfect the MPod – Quad Hemo with a gauze pad moistened with diluted
alcohol.
3. Dry thoroughly with a lint-free cloth.
SpO
2
Clean reusable SpO
solution. To disinfect sensors, wipe using a cloth moistened with a 70% alcohol
solution. Dry thoroughly with a lint-free cloth before applying to patient.
Reusable SpO
Refer to the cleaning instructions and recommendations provided with the sensor.
Infinity Masimo SET SpO
1. To clean the Masimo SET SpO
2. Clean the pod with a gauze moistened in enzymatic detergent or a solution of
green tinctured soap and water. Dry thoroughly with a lint-free cloth.
CAUTION:
29-6
D
ISINFECTING
sensors by wiping them with a gauze pad moistened with a soap
2
Sensor
2
Do not use organic solvents.
Do not sterilize by steam, heat, radiation or ETO.
Do not use sharp objects.
Make sure that NO liquid enters the pod.
D
/D
ELTA
Pod
2
pod disconnect it from the monitor.
2
XL/K
ELTA
APPA
VF8