Dräger Infinity Delta Instructions For Use Manual page 396

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

23 S
® F
CIO
OUR
NOTE:
NOTE: In case of a power outage, the Scio module shuts off and when power is
returned, the module comes back on.
NOTE: Dräger pods and accessories that come into contact with the patient are all
free of latex.
23-6
M
ODULES
Breathing gases are sampled at a flow rate of 200 ±20 or 150 ±20 mL/min as
specified. The sample flow rate is specified on the back panel. If the flow rate is
not specified on the back panel, the module has a flow rate of 150 ±20 mL/min.
The Scio module purges and zeroes itself approximately once every two hours.
The typical purging/zeroing cycle lasts no longer than 25 seconds. Waveforms
flatline and parameter box values blank from the screen during this cycle. The
message Multigas Zero in Progress appears in the message box. (An extended
zeroing cycle may be performed after the unit is initially powered up).