Dräger Infinity Delta Quick Reference Manual
Hide thumbs Also See for Infinity Delta:

Advertisement

Quick Reference Guide
Infinity Delta / Delta XL
/ Kappa
This Infinity Quick Reference Guide is
not a replacement or substitute for
the Instructions for Use (IFU). It is for
informational purposes
Software Level: VF9
only.

Advertisement

Table of Contents
loading

Summary of Contents for Dräger Infinity Delta

  • Page 1 Quick Reference Guide This Infinity Quick Reference Guide is not a replacement or substitute for the Instructions for Use (IFU). It is for informational purposes only. Infinity Delta / Delta XL Software Level: VF9 / Kappa...
  • Page 2: Table Of Contents

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Content Electrode & Lead Placement Docking, Transport, & Power Fixed Keys Patient Category, Admit, Discharge, Standby, Pri- 10-11 vacy, & Transferring Data Alarms Priorities Alarm Pause Alarm Limits Table 14-15 Fast Access Fixed Key Event Recall &...
  • Page 3: Electrode & Lead Placement

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Electrode & Lead Placement DO’S & DO NOT’S Do place the electrode on a clean and dry area of skin. Do utilize soap, water, and dry skin thor- oughly. Do remove hair if necessary for proper elec- trode placement.
  • Page 4 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Electrode Colors* AHA/US Electrode Location Yellow LA Black Left arm Green LL Red Left leg RA White Right arm N Black RL Green Right leg (neutral) White V Brown Moveable chest C+ Gray & White V+ Gray &...
  • Page 5 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Standard 3 Lead Configuration Pacemaker Configuration 5 electrodes Warning: Make sure the pacemaker detection is turned OFF for patients without pacemakers and that it is ON for patients with pacemakers. Al- ways keep pacemaker patients under close surveillance.
  • Page 6 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE 12-Lead Configuration (ten electrodes) Respiratory lead placement for infants THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION © 2014 Draeger Medical, Inc.
  • Page 7: Docking, Transport & Power

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Docking, Transport & Power PICK AND GO TRANSPORT (Delta/Delta XL only) The Pick and Go patient transport system allows the monitor to travel with the patient to different care stations within the hospital.
  • Page 8 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE BATTERY POWER The Delta/Delta XL has an internal lithium ion battery, which can power the monitor while on transport for approximately 240 minutes when new. When the monitor is re-docked the battery will recharge.
  • Page 9 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Press the knob (or click) to confirm THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION © 2014 Draeger Medical, Inc.
  • Page 10: Fixed Keys

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Fixed Keys Function ALARM SILENCE Silences the active alarm tone for 2 minutes or cancels the Alarm Silence *Note: Silences the active alarm tone for 1 mi- nute when Backwards Compatibility is Enabled,...
  • Page 11: Patient Category, Admit, Discharge, Standby, Pri

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Patient Category, Admit, Discharge, Standby, Privacy, Transferring Data PATIENT CATEGORY: ADULT, PEDIATRIC, AND NEONATAL Press Menu fixed key. Use Rotary knob to highlight Patient Setup, click. Use Rotary knob to highlight Patient Category, click.
  • Page 12 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE TO ACCESS PRIVACY SCREEN Privacy screen is displayed at the bedside with all Monitoring and Alarms taking place at the Central Station. Press Menu fixed key. Highlight Privacy, click. To release, press Main Screen key.
  • Page 13: Alarms Priorities

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Alarm Priorities HIGH-PRIORITY ALARMS Report conditions that can be life threatening and require im- mediate intervention; for example, asystole. The alarming parameter’s box and alarm message field have red background, and audible alarms*sound when active.
  • Page 14: Alarm Pause

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Alarm Pause The monitor issues alarms for parameters that are set to ON in the Alarm Limits table. It is not a prerequisite for the parameter to be on display or connected for a parameter to alarm. Turning a param- eter’s alarm function Off prevents the parameter from triggering an...
  • Page 15: Alarm Limits Table

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Alarm Limits Table The alarm limits table allows you to modify the alarm limits of mul- tiple parameters in a single location. Press the Alarm Limits fixed key - The Alarm Limits table appears...
  • Page 16 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE CONFIGURE PARAMETER ALARM LIMITS Press Alarm Limits fixed key Use Rotary Knob to highlight the parameter you wish to con- figure, i.e. NBP, click Use Rotary knob to adjust Alarms ON/OFF, click Use Rotary knob to adjust Upper limit, or Lower Limit, click Use Rotary knob to adjust Archive setting Off –...
  • Page 17 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Fast Access Key-Data EVENT RECALL/PRINT STORED EVENTS Events are saved to Event Recall from a parameter alarms, arrhyth- mia events, and Marked Events. Press FAST ACCESS fixed key. Use Rotary knob to highlight Event Recall, click.
  • Page 18 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE TRENDS Press FAST ACCESS fixed key. Use Rotary knob to highlight Trend Graph or Trend Table, click. Use Rotary knob to highlight interval time, click. Use Rotary knob to set desired interval, click.
  • Page 19: Screen Setup/Auto Display

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Menu Item Description Setting Options Table Interval Determines time interval for tabular re- 1, 5, 15, 30, 60 port. Corresponds to the interval setting minutes at bottom of the trend table. Screen Setup/Auto Display...
  • Page 20: Parameter Boxes

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Use Rotary knob to move parameter to new position, click. Screen will refresh with new parameter priority. Parameter Boxes You can access parameter setup menus by scrolling a blue box through the parameter boxes using the rotary knob and selecting the parameter you wish to configure.
  • Page 21: Respiratory Monitoring

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE TO DISPLAY ALL MONITORED LEADS Use Rotary knob to highlight HR parameter box, click. Use Rotary knob to select Show All Leads, click. Use Rotary knob to highlight REPORT, click. A non-diagnostic ECG Report will be sent to the networked laser printer.
  • Page 22 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE optimize breath count. Manual The monitor detects breath based on the size of the waveform. After turning on Manual mode adjust waveform size. Intended for adult or pediatric patients whose breathing patterns show excessive variation; or for neonates whose breathing rhythms tend to be irregular, and whose signals may not be reliably detected in Auto mode.
  • Page 23: Temperature

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Learns the patient’s respiration pattern when the Mode selection is set to Auto Use Rotary knob to highlight Respirations parameter box, click Use Rotary knob to highlight Relearn, click Note: Initiate a relearn if electrodes have been repositioned or if the patient's breathing pattern changes.
  • Page 24: Advanced Parameters

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE TO ADJUST SpO2 SETTINGS Use Rotary knob to highlight SpO2 parameter box, click. Use Rotary knob to highlight Waveform Size, adjust, and click. Use Rotary knob to highlight Pulse Tone Source, adjust, click.
  • Page 25: Pulmonary Artery Pressures

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE TO ZERO TRANSDUCERS Y-CABLE Use Rotary knob to highlight the Pressure Parameter box, click. Use Rotary knob to highlight Zero, click. TO CHANGE LABELS Use Rotary knob to highlight the Pressure Parameter box, click.
  • Page 26: Cardiac Output

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Use Rotary knob to adjust Scale (5, 10, 15, 20, 25, 30, 35, 40, 50, 75, 100, 125, 150, 175, 200, 225, 250, or 300 mmHg), click. Use Rotary knob to adjust Sweep Speed (6.25, 12.5, 25, or 50 mm/s), click.
  • Page 27: Hemo/Oxy/Ventilation Calculations

    INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE structions for Use manual, or reference guide from cathe- ter manufacturer based on the catheter size, and the vol- ume, and temperature of injectate. TO PERFORM A CARDIAC OUTPUT MEASUREMENT Choose Auto or Manual Injection o In Auto Mode, press the CO Start key on the Hemopod.
  • Page 28 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION © 2014 Draeger Medical, Inc.
  • Page 29 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Microstream etCO2 Pod Attach the connecting cable from the Pod to the Aux port on the monitor. Attach the appropriate sampling line to the Pod. Move cover of CO2 port to side; insert the sampling line connecter into CO2 port;...
  • Page 30 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE CONNECTING ETCO2 MODULES AND PODS Ports and outlets are clearly marked on the front of the etCO2 module and pod. Use these labels as a guide when connecting the pod or module to the monitor and peripherals.
  • Page 31 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Select a sidestream airway adapter. Make sure the windows are clean and dry. Use the sidestream sampling tubing to con- nect the airway adapter to the input connector on the etCO2 module or pod.
  • Page 32 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Insert the adapter in a vertical position between the elbow and ventilator circuit “Y, and position sensor cable away from patient. Confirm proper connections to ventilator and patient. Access the etCO2 menu and confirm correct settings. Adjust...
  • Page 33 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE Intentionally left blank. THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION © 2014 Draeger Medical, Inc.
  • Page 34 INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE HEADQUARTERS Dräegerwerk AG&Co. KGaA Moslinger Allee 53-55 23558 Lübeck, Germany www.draeger.com CANADA Draeger Medical Canada Inc, 2425 Skymark Ave Unit1 Mississauga, ON L4W 4Y6 +1 905 212 6600 Toll-free +1 866 343 2273...