Dräger Infinity Delta Instructions For Use Manual page 538

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

B T
D
ECHNICAL
Delay:
Pacemaker pulses:
Output range
Signal gain
ECG bandwidth
Pacemaker pulses
QRS Sync Output
Delay:
Output high (QRS
detected):
Output low (no QRS):
Pacemaker pulses:
User Interface
Controls:
Alarms:
Screen
Resolution:
Sweep speed:
Trend storage
Data storage:
Data resolution:
Trend graphs:
Trend tables:
B-12
ATA
≤25 ms
Included
–5 V to +5 V
ECG: X 1000
ART: X 10 mV/mmHg
0.5 to 20 Hz (ESU filter mode)
0.5 to 40 Hz (other filter modes)
Amplitude: 5 V (nominal)
Duration: 4 ms
Rise time: <18 ms
Fall time: <20 ms
≤35 ms
Duration: 50 ms
Amplitude: +7.1 V ±5 % (nominal) – rear panel sync. connector
Amplitude: +12 V ±5 % (nominal) – analog/sync. connectors
Source impedance: 5100 Ω
Rise time: <7 µs
Fall time: <1 µs
<0.8 V @30 mA sink current
Not included
Fixed keys and rotary knob
3 levels: High, Medium, Low
800 x 600
6.25, 12.5, 25 and 50 mm/s ±10 % (accuracy only guaranteed for a
15" display)
24 hours of trended parameter information
30 second sampling
1, 2, 4, 8, 12, and 24 hour display formats
1, 5, 15, 30, and 60-minute display formats