Thermodilution Setup Menu
Menu Item
Catheter type
NOTE: Other is
displayed if the
catheter that is
connected requires
manual entry of an
ACC value.
Selecting Other
displays a key pad
for manual entry.
Catheter Position
Art. Catheter
Const. (ACC)
Manual CVP
VF8
D
Description
Read-only field. Displays the
catheter type in use.
Selects the position of the catheter.
NOTE: The selection Radial artery
is only visible when the catheter
PV2014L50 is connected.
Field ghosted when catheters with
automatic ACC detection are
connected. This field will be
unghosted and available to catheters
(other than Catheter Type field)
requiring manual entry of catheter
constant values (available range is
0-999). Selecting this key displays a
keypad for manual entry.
Allows you to enable/disable manual
CVP Value menu.
NOTE: If Manual CVP is Enabled,
CVP Value selection is ghosted
(unavailable).
/D
XL/K
ELTA
ELTA
P
CCO P
I
Available Settings
PV2013L07, PV2014L08,
PV2014L16, PV2014L22,
PV2015L20, PV2014L50, Other
Femoral artery, Axillary artery,
Brachial artery, Radial artery
NOTE: When a patient is initially
admitted, this field displays '---' until
the catheter position is selected.
CAUTION: Confirm that the
selected catheter position is
accurate before starting a
measurement. Changing the
catheter position after starting a
measurement, has no effect on the
current measurements. To correct
the catheter position, discharge and
readmit the patient.
Not applicable
Disabled (default), Enabled
APPA
S
ARAMETER
ETUP
26-21