Menu Access

The FiO
menu displays the date and time of 1-point and 2-point calibrations (see
2
below). The menu item
time of the last successful calibration, either 1-point or 2-point.
To open the FiO
Click on the
or
1. Press the
2. Click on
3. Click on
4. Click on
Calibration
Every time a sensor is connected to the NeoMed pod, you must calibrate the monitor
to the sensor. The monitor does not display FiO
There are two types of calibration. 1-point calibration measures the oxygen in room
air, typically 21 %, and calibrates the monitor to that measurement. A 2-point
calibration uses two measurements, room air and 100 % oxygen, to calibrate the
monitor. A 2-point calibration provides more accurate FiO
monitor is calibrated to two different measurements.
1-point calibration should be performed daily. A 2-point calibration should be
performed weekly. You should also calibrate the monitor under the following
circumstances:
Periodically, to verify the correct functioning of the sensor
Daily, if you are monitoring a patient's FiO
When you suspect that sensor characteristics have changed
When the accuracy of the monitor is in question
When there is a change in humidity or barometric pressure of the monitoring
site
22-4
I
NSPIRED
is informational only. It displays the date and
Last O2 Cal
menu
2
parameter box on the Main screen
FiO
2
fixed key.
Menu
.
Patient Setup
.
Parameters
.
FiO2
D
/D
ELTA
ELTA
O2)
MONITORING
values until it is calibrated.
2
on a daily basis
2
XL/K
APPA
monitoring because the
2
VF8