26 P
C
ULSE
ONTOUR
The transpulmonary thermodilution PiCCO parameters are listed in the following
table.
Transpulmonary Thermodilution parameters
Label
Parameter
p-CO
Cardiac Output - Transpulmonary
thermodilution cardiac output serves as the
basic parameter for calculation of various
blood volumes.
p-CI
Cardiac Index - Cardiac output normalized to
BSA (Body Surface Area).
GEDV
Global End Diastolic Volume - The sum of all
end-diastolic volumes of the atria and the
ventricles. GEDV is equivalent to preload
volume of the total heart.
GEDVI
Global End Diastolic Volume Index
GEDVI = GEDV / PBSA (Predicted Body
Surface Area)
EVLW
Extravascular Lung Water - correlates to
extravascular thermal volume in the lungs and
is determined by the mean transit time
method.
EVLWI
Extravascular Lung Water Index -
Extravascular Lung Water normalized to body
weight.
EVLWI = EVLW / PBW (Predicted Body
Weight)
GEF
Global Ejection Fraction - equal to 4 times the
stroke volume divided by the global end
diastolic volume (GEDV).
PVPI
Pulmonary Vascular Permeability Index -
Indicator of capillary leakage.
CFI
Cardiac Function Index - CFI = CI / GEDVI
ITBV
Intrathoracic Blood Volume.
ITBVI
Intrathoracic Blood Volume Index.
ITBVI = ITBV / PBSA (Predicted Body Surface
Area)
NOTE:
• Normal values reflect clinical experience and may vary from patient to patient. They are without
guarantee.
• The prefix of "p-" before SV, SVI, SVR, SVRI is an indication that these parameters are derived
by the PiCCO pod.
26-4
C
O
ARDIAC
UTPUT
D
/D
ELTA
ELTA
(P
CCO)
I
MONITORING
Measurement
Range
0.25 to 25 L/min
0.10 to
2
15.00 L/min/m
40 to 4800 ml
80 to 2400 ml/m
10 to 5000 ml
0 to 50 ml/kg
1 to 99%
0.1 to 10
1.0 to 15.0 L/min
50 to 6000 ml
100 to
2
3000 ml/m
XL/K
APPA
Normal Value
Not applicable
2
3.0 - 5.0 L/min/m
Not applicable
2
2
680 to 800 ml/m
Not applicable
3.0 to 7.0 ml/kg
25 to 35%
1.0 to 3.0
4.5 - 6.5 L/min
Not applicable
2
850 - 1000 ml/m
VF8