Resp Setup Menu - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

12 R
ESPIRATION

RESP Setup Menu

All impedance respiration functions are controlled from the RESP setup menu, which
you can open in one of two ways:
To access the RESP setup menu
Click on the
or
1. Press the
2. Click on
3. Click on
appears.
4. Scroll to
Quick Reference Table
The Respiration Setup Menu
Menu Item
Description
RESP Lead
Determines respiration lead
Mode
Determines processing mode for
breath-related impedance changes
WARNING: If the respiration
waveform size is set too low,
shallow breaths may not be
counted. If it is set too high,
cardiac artifact will be counted as
breaths. Therefore, always use
the respiration marker to verify
breath detection at the desired
amplitude.
Size
Auto mode – Adjusts the waveform
size.
Manual mode – Adjusts the
waveform size and the breath-
detection threshold.
12-6
parameter box on the main screen
RESP
fixed key on the front of the monitor.
Menu
. A list of available patient setup functions appears.
Patient Setup
in the second column. A list of available parameters
Parameters
and click to display the RESP setup menu.
RESP
Respiration Setup
D
/D
ELTA
ELTA
Settings
I, II
• Auto – Optimal breath-detection threshold
calculated at the beginning of RESP
monitoring. Intended for patients with
regular breathing patterns.
• Manual – The monitor does not set a
breath-detection threshold at the
beginning of RESP monitoring. Instead,
the adjustments you make to the
waveform Size (see the Size setting,
below) set the monitor's breath detection
sensitivity. Intended for adult or pediatric
patients whose breathing patterns show
excessive variation; or for neonates whose
breathing rhythms tend to be irregular, and
whose respiration signals may not be
reliably detected in Auto mode.
• 5 - 100 % (in 5 % increments)
XL/K
APPA
VF8