Sampling Methods

Mainstream
For mainstream detection, the sensor is located within the patient's airway or
breathing circuit, allowing you to monitor individual breathing cycles of the intubated
patient. This method is appropriate for neonatal as well as adult and pediatric patients.
Mainstream Monitoring Setup
1. Click on etCO
etCO
2. Click on
3. Click on
4. Select a mainstream airway adapter.
Make sure the windows are clean and
dry. Clean or replace the adapter if
necessary.
5. Align the marks on bottom of the adapter
with the sensor.
6. Snap firmly into place.
7. Insert the adapter in a vertical position
between the elbow and ventilator circuit "Y.
8. Position sensor cable away from patient.
9. Confirm proper connection on the etCO
19-6
-T
CO2)
IDAL
During nebulization or suction for intubated patients, remove the sampling line
luer connector from the monitor to avoid moisture buildup and sampling line
occlusion.
Replace the sampling line according to hospital protocol or when a blockage is
indicated by the device. Excessive patient secretions or a build-up of liquids in
the airway tubing may occlude the sampling line, requiring more frequent
replacement.