Dräger Infinity Delta Instructions For Use Manual page 343

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Function
Description
Respiratory
Sets waveform sweep speed on screen display. • 6.25 mm/s
Sweep Speed
Averaging
Sets interval for CO
NOTE: The monitor reports the maximum value
of etCO
Atm. Pressure
Determines automatic or manual setup for
Mode
atmospheric pressure compensation
NOTE: Atmospheric pressure compensation
should be set by qualified personnel only.
Consult with your hospital's technical personnel/
DrägerService before changing the Atm.
Pressure Mode setting to Manual.
Atm. Pressure
This is the current atmospheric pressure.
This function is ghosted if you have selected
Auto under Atm. Pressure Mode.
Measurement
Configures the monitor for mainstream or
Mode
sidestream monitoring
Respiration Filter
Filters signal artifact caused by cardiogenic
(module only)
oscillations or Baines rebreathing bumps, which
could cause the monitor to display an
erroneously high respiration rate (RRc)
Apnea Alarm
Displays the Apnea Alarm menu.
etCO2 Alarm
Accesses the Alarm Limits table, where you can
set upper and lower alarm thresholds.See
chapter 5, Alarms, for information about setting
and displaying alarm limits.
VF8
D
measurements
2
during the specified sampling interval.
2
/D
XL/K
ELTA
ELTA
Settings
• 12.5 mm/s
• 25 mm/s
• 50 mm/s
• Breath (end-expiration
point)
• 10 s
• 20 s
• Instantaneous (pod only
setting) – Represents the
highest CO
one-second interval
• Auto
• Manual
• 540 to 800 mmHg (in
increments of 5)
• Main
• Side
• Normal – Disables the
filter
Special –
filtering at the expense of
a certain degree of
monitor responsiveness
• RRc Apnea Time
• Apnea Archive
Not applicable
APPA
CO2 S
ET
ETUP
value per
2
Enables this
19-13