Dräger Infinity Delta Instructions For Use Manual page 309

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Message
SpO
Searching
2
SpO
Unplugged
2
SpO
: Pod Failure
2
SpO
: Check/
2
replace Sensor
SPO
: non-Nellcor
2
compatible sensor
Duplicate Device
Connected
VF8
D
Possible Cause
The monitor has not been able to
compute a valid measurement and is
searching for a pulse.
The sensor cable not connected to
the MultiMed pod.
Pod hardware or software failure.
Defective/faulty sensor
A non-Nellcor compatible sensor or
SpO
pod is connected to a monitor
2
that is configured for Nellcor.
A MultiMed pod (with SpO
and an Infinity Nellcor Oximax pod
are connected simultaneously.
/D
XL/K
ELTA
ELTA
Suggested Action
• Make sure the SpO
properly attached to the patient at a
site free of patient motion and all
cables are properly connected.
• Make sure the correct sensor type
and/or site is used.
• Contact DrägerService.
• Make sure the cables are securely
connected.
• Check for a defective sensor.
• Disconnect/reconnect the pod if
problem persists.
• Disconnect/reconnect the pod and/
or replace with a good sensor.
• Contact the hospital's technical
personnel (SpO
locked option).
• Replace with a Nellcor-compatible
sensor.
• Contact the hospital's technical
personnel (SpO
locked option).
• Replace with a Nellcor-compatible
sensor.
• Contact the hospital's technical
personnel (SpO
locked option).
sensor)
Disconnect the duplicate device.
2
APPA
S
M
TATUS
ESSAGES
sensor is
2
compatibility is a
2
compatibility is a
2
compatibility is a
2
17-19