Message
SpO
Regulation
2
Error
SpO
Unplugged
2
SpO
: non-Masimo
2
compatible sensor
Duplicate Device
Connected
VF8
D
Possible Cause
• Inconsistent light level
• Excess ambient light
The sensor cable is not
connected to the MultiMed pod.
A non-Masimo compatible
sensor or SpO
pod is
2
connected to a monitor that is
configured for Masimo.
A MultiMed pod (with SpO
sensor) and an Infinity Masimo
SET pod are connected
simultaneously.
/D
XL/K
ELTA
ELTA
Suggested Action
• Make sure the SpO
attached to the patient.
• Remove or cover any external sources
of light entering the sensor.
• Contact DrägerService.
• Make sure the cables are securely
connected.
• Check for a defective sensor.
• Replace with a Masimo-compatible
sensor.
• Contact the biomedical personnel
(SpO
compatibility is a locked option).
2
Disconnect the duplicate device.
2
APPA
S
M
TATUS
ESSAGES
sensor is properly
2
17-17