Dräger Infinity Delta Instructions For Use Manual page 303

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Message
SpO
Searching
2
SpO
Unplugged
2
SpO
: Pod Failure
2
SpO
: Check/
2
replace Sensor
SpO
: Connection
2
Not Supported
VF8
D
Possible Cause
The monitor has not been able to
compute a valid measurement and is
searching for a pulse.
The sensor cable is not connected to
the MultiMed pod.
Pod hardware or software failure
Defective/faulty sensor
The Masimo SET pod is connected.
/D
XL/K
ELTA
ELTA
Suggested Action
• Make sure the SpO
properly attached to the patient at a
site free of patient motion and all
cables are properly connected.
• Make sure the correct sensor type
and/or site is used.
• Contact Dräger DrägerService.
• Make sure the cables are securely
connected.
• Check for a defective sensor.
• Disconnect/reconnect the pod, if the
problem persists.
• Disconnect/reconnect the pod and/
or replace it with a good sensor.
• Contact the hospital's technical
personnel (SpO2 compatibility is a
locked option).
• Replace with a Nellcor-compatible
sensor.
• Contact the hospital's technical
personnel (SpO2 compatibility is a
locked option).
Disconnect the Masimo SET pod and
connect a Nellcor Oximax pod.
APPA
S
M
TATUS
ESSAGES
sensor is
2
17-13