5. Fill the injectate temperature sensor housing, enclosed in the pressure
monitoring kit and connect it to the distal lumen of the central venous line.
6. Connect the p-CO thermistor cable to the p-CO intermediate cable.
7. Plug the p-CO intermediate cable into the port labeled "CO" on the PiCCO
pod.
8. Connect the "Injectate temperature sensor housing" to the p-CO thermistor
cable.
9. Slide the ART blood pressure transducer into a slot on the front of the PiCCO
pod. Make sure the slot's pressure label is "ART".
10. Slide the Central Venous Pressure transducer into another slot. Make sure this
slot pressure label is "CVP".
11. Connect the PiCCO catheter with the blood pressure transducer.
12. Use the 10-pin cable to connect the blood pressure transducer to the Delta/
Delta XL/Kappa.
13. If you have not already done so, enter the patient specific input parameters
(height and weight) in the Patient Admit menu (see page 4-2).
NOTE: PiCCO measurements are only possible after the arterial pressure transducer
is zeroed and the PiCCO pod is calibrated. For accurate measurements, calibrate the
PiCCO pod within five minutes of zeroing the pressure.