Labeling Pressure Channels - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

14 I
B
NVASIVE
IBP Setup Menus
Menu Item
Manometer Cal
Pressure Overlap
Common Scale
Large Mean
Wedge Start
(PA setup menu
only)
Min Scale
(ICP, ICP2, ICP3,
ICP4 setup menus
only)
Pressure Labels
<IBP Parameter>
Alarms

Labeling Pressure Channels

The pressure label determines how a signal is analyzed and reported to the monitor.
For detailed information about the various types of pressure labels, see page 14-2.
When you assign a new label to a pressure channel, the monitor clears the parameters
and conditions set for the previous label (including alarms and waveform scales) and
replaces them with settings for the new label. Trends are stored according to the
assigned label. Zero, Cal Factor, and Cal Date & Time settings are associated with
the pressure channel and are kept even during a label change.
The
Pressure Labels
monitor assigns an automatic pressure label (
Signal sources are displayed in rows [1-3], with Row 1 representing data received
from the HemoMed pod or Y-cable, and Rows 2 and 3 representing data from the
Hemo2, Hemo4, MPod – Quad Hemo or PiCCO pod, either of which may be
14-14
P
LOOD
RESSURE
Description
Allows you to enter the manometer or
simulator reading and start calibration
(see page 14-5).
Allows you to view up to four IBP
parameters on a single baseline.
Sets the waveforms to one scale.
Increases size of all IBP mean values.
Starts a wedge pressure measurement
Same function as the Wedge fixed key
on a hemodynamic pod (see page
14-18),
NOTE: Not available on the Picco pod.
Sets the ICP waveform scale to +/- 25
mmHg (+/- 3.3 kPa).
NOTE: When the Min Scale setting is
set to ON, the Scale selections are
unavailable.
Displays the Pressure Label screen
(see page 14-14).
Displays alarms for parameter and
associated variable on Alarm Limits
table.
screen can display up to twelve IBP sources in a 3
D
/D
ELTA
Available Settings
10 – 300
• ON
• OFF
• OFF, 5, 10, 15, 20, 25, 30, 35,
40, 50, 75, 100, 125, 150, 175,
200, 225, 250, and 300
• ON
• OFF
Not applicable
• ON
• OFF
Not applicable
Not applicable
[
-
][
-
]) to each box.
P
1
3
a
d
XL/K
ELTA
APPA
4 matrix. The
X
VF8