Dräger Infinity Delta Instructions For Use Manual page 165

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Definition and/or
Message
Possible Cause
ECG Artifact
• Patient movement, shivering,
tremors
• Bad electrode contact
• Excessive signal noise
• Interference from auxiliary
equipment
Collecting ECG
The monitor has begun a 10-
Waveforms
second capture of Rest ECG
waveforms.
ECG Leads
• The QRS processing leads are
Invalid
invalid for > 10 sec.
• Bad electrode contact or faulty
lead set
• Unplugged lead set
• The wrong cable type was
selected on the ECG Lead setup
menu (see page 8-20).
ECG Report
The Infinity CentralStation is
Server Busy-Try
currently processing a report.
Later
ECG
An ECG hardware failure was
H/W Failure
detected by an 1 mV test at
startup.
MultiMed Pod
The MultiMed pod is not connected
Disconnected
to the monitor during 3, 5, or 6-lead
monitoring.
VF8
D
/D
XL/K
ELTA
ELTA
Suggested Action
• Calm the patient.
• Check the electrodes and reapply, if
necessary.
• Make sure the patient's skin is properly
prepared.
• Isolate the patient from auxiliary
equipment, if possible.
Instruct the patient to lie still.
• Inspect and replace defective cables and
wires.
• Reapply gel on disposable electrode(s).
• If a lead or electrode cannot be replaced,
select another ECG lead for processing.
• Verify that the number of leads selected in
the ECG Lead Setup menu matches the
applied lead set (see page 8-20).
Wait a few minutes, then try again.
Contact the hospital's technical personnel
or DrägerService.
• Check the cables and connection; replace
cables, if necessary.
• If the MultiMed pod is not in use, press the
Alarm Silence fixed key.
APPA
S
M
TATUS
ESSAGES
8-25