Display Features; Tpo2/Co2 Setup - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

18 T
RANSCUTANEOUS

Display Features

The monitor displays tpO
In the tpO
In the trend display (see chapter 6, Trends)
On the oxycardiorespirogram (see page 12-9)
When two tpO
screen. The labels of the second pod are marked by an asterisk (*), as shown below:
tpO
/CO
2
2
High sensor temperatures on the skin's surface offer the greatest degree of accuracy
and can quickly detect a change in the patient's condition. However, the heat from the
sensor may cause severe burns, especially in neonates. On the tpO
you can restrict the application time of the sensor and establish a temperature set point
(the target temperature for heating the sensor during calibration).
The tpO
/CO
2
measurement. The tpO
current barometric pressure, which changes with altitude and weather conditions. The
monitor allows you either to rely on the barometric pressure values generated
automatically by the pod or manually enter pressure values obtained with an external
barometer. Readings are displayed on the tpO
See "Precautions" on page 18-3 for other important safety information.
18-10
B
LOOD
/CO
values and trends as follows:
2
2
/CO
parameter box on the Main screen
2
2
/CO
pods are connected, each displays a parameter box on the main
2
2
Setup
pod's settings for altitude and barometric pressure influence the
2
/CO
pod contains an internal pressure sensor that measures
2
2
D
/D
ELTA
G
M
AS
ONITORING
/CO
setup menu.
2
2
XL/K
ELTA
APPA
/CO
setup menu,
2
2
VF8