Quick Reference Table -- Nmt Monitoring Settings - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

24 N
EUROMUSCULAR

Quick Reference Table -- NMT Monitoring Settings

Menu Item
Start NMT
Stop NMT
Mode
Measurement Interval
Pulse Width *
Stim Current Mode
Stim Current *
*Caution: Changing the stimulation current or pulse width invalidates the stored reference value.
The pod recalculates the reference twitch at the next measurement, based on the existing level of
patient relaxation, which could cause misleading NMT measurements.
24-8
T
RANSMISSION
Description
Starts NMT
measurements.
Stops NMT measurements. Not applicable
Allows you to select
measurement mode.
Allows you to select time
interval for automatic NMT
measurement cycles (in
single and TOF modes).
Allows you to select
duration for stimulation
pulse.
Allows you to select
stimulation current mode.
• In auto mode, the pod
establishes a
supramaximal current
during the first
measurement and uses it
for subsequent
measurements.
• In manual mode, the user
selects a stimulation
current.
Allows you to select
stimulation current
(read-only field in Auto
mode)
D
/D
ELTA
ELTA
(NMT)
MONITORING
Available Settings
Not applicable
• Single
• TOF (default)
• PTC
None, 1s, 10s, 20s (default), 1min,
5min, 15min, 30min
NOTE: in PTC mode, you cannot select
a measurement interval; in TOF mode,
1s is not available.
100 µs, 200 µs (default), 300 µs
• Auto (default)
• Manual
• Auto mode: none
(displays the stimulation current value)
• Manual mode: 5 mA (default) - 60 mA,
(in increments of 5 mA)
XL/K
APPA
VF8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents