APPENDIX 2: NARKOMED 3 ALARM MESSAGES

CLASS
ALARM MESSAGE
W
% OXYGEN LOW
W
APNEA - PRES
W
APNEA - VOL
W
APNEA - CO2
W
VENT PRES HI
W
SUB ATM PRES
W
NO OXI PULSE
W
OXI PULSE LO
W
OXI SAT LOW
W
% HAL HIGH
W
% ENF HIGH
W
% ISO HIGH
C
LO O
C
APNEA - PRES
84
% O
Apnea for 30 seconds
Apnea for 30 seconds
Apnea for 30 seconds
Pressure > high limit
Pressure < -10 cm H
No pulse for 10 seconds
Pulse < low limit
Saturation < low limit
Halothane % > high limit
Enflurane % > high limit
Isoflurane % > high limit
SUPPLY
O
supply pressure < 37 psi
2
2
Apnea for 15 seconds
RETURN TO CD-ROM TABLE OF CONTENTS
CONDITION
< low limit setting
2
O
2
AUDIO
CONT
CONT
CONT
CONT
CONT
CONT
CONT
CONT
CONT
CONT
CONT
CONT
INTERMIT
INTERMIT