Dräger Infinity Delta Instructions For Use Manual page 404

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

23 S
® F
CIO
OUR
The etCO
* Setup Menu
2
Menu Item
Description
Auto Zero
Delays automatic zeroing for 5 minutes for
Delay
uninterrupted monitoring.
WARNING: Delaying the auto zero may impact
accuracy.
NOTE: Gas sensors in the Scio module are
automatically zeroed and calibrated against room
air. Parameter values disappear during zeroing. One
minute before automatic zeroing, the monitor sounds
an attention tone and displays the message Auto
zero in <1 minute.
etCO2*
Accesses etCO
Alarms
WARNING: etCO
after turning on the monitor or discharging a patient.
23-14
M
ODULES
* alarms in Alarm Limits table (see chapter 5)
2
alarms do not activate until the first breath is detected
2
D
/D
ELTA
ELTA
Settings
Not applicable
XL/K
APPA
VF8