To measure C.O. in manual mode
1. Press the
the C.O. setup menu. The READY message appears for 30 seconds or until a
blood temperature drop is detected.
2. Immediately inject the saline solution and wait for the monitor to calculate a
C.O. value. The monitor begins to calculate a C.O. value as soon as it detects
a blood temperature drop.
If the monitor fails to detect the temperature drop caused by the injectate, the
waveform disappears after 30 seconds. An attention tone sounds, an error
message appears in the local message area, and three asterisks (
the Save AVG field. Repeat steps 1 and 2 for additional measurements.
NOTE:
C.O. Start
initiate a C.O. measurement, the
is ghosted until a value is reported.
VF8
D
fixed key on the hemodynamic pod or select
C.O. Start
is only available as a menu item in manual mode. As soon as you
C.O. Start
/D
XL/K
ELTA
ELTA
C.O. M
EASUREMENT
key is disabled and the
APPA
P
ROCEDURES
on
C.O. Start
appear in
* * *)
menu item
C.O. Start
15-11