Non-Invasive Blood Pressure (NBP)
Parameter display:
Measurement method:
Modes of operation:
Interval times:
Measurement range (heart
rate)
Pressure measurement
range
(Adult - 270 mmHg):
Pressure measurement
range
(Pediatric - 180 mmHg):
Pressure measurement
range
(Neonatal - 140 mmHg):
Connections:
Default inflation pressure:
Inflation pressure after a
valid measurement
(±10 mmHg):
Inflation pressure
after an alarm:
Maximum inflation
pressure:
Minimum inflation
pressure:
Maximum measurement
time:
Maximum measurement
time including a retry:
Software safety cut-off:
Hardware safety cut-off:
Static cuff accuracy:
Calibration range:
VF8
D
Systolic, Diastolic, Mean
Oscillometric technique
Manual (single measurement), Continuous (5 minutes), or Interval
1, 2, 2.5, 3, 5, 10, 15, 20, 25, 30, 45, 60, 120, and 240 min
30 to 240 bpm
Systolic NBP: 30 to 250 mmHg
Mean NBP:
20 to 230 mmHg
Diastolic NBP: 10 to 210 mmHg
Systolic NBP: 30 to 170 mmHg
Mean NBP:
20 to 150 mmHg
Diastolic NBP: 10 to 130 mmHg
Systolic NBP: 30 to 130 mmHg
Mean NBP:
20 to 110 mmHg
Diastolic NBP: 10 to 100 mmHg
Quick-release hose connector with single airway
Adult (270):
160 mmHg ±10mmHg
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Adult (270):
Pediatric (180): Previous NBP
Neonatal (140): Previous NBP
Adult (270):
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Adult (270):
Pediatric (180): 180 mmHg ±10 mmHg
Neonatal (140): 142 mmHg ±10 mmHg
Adult (270):
Pediatric (180): 90 mmHg ±10 mmHg
Neonatal (140): 70 mmHg ±10 mmHg
Adult (270):
Pediatric (180): 2 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Adult (270):
Pediatric (180): 3 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Adult (270):
Pediatric (180): 215 ±3 mmHg
Neonatal (140): 153 ±3 mmHg
Adult (270):
Pediatric (180): 300 ±30 mmHg
Neonatal (140): 157 ±8 mmHg
±3 mmHg
Adult and Pediatric:10 to 260 mmHg ±3 mmHg
Neonatal:
/D
XL/K
ELTA
ELTA
M
ONITORING
Previous NBP
+ 25 mmHg
SYS
+ 25 mmHg
SYS
+ 30 mmHg
SYS
160 mmHg ±10 mmHg
265 mmHg ±5 mmHg
110 mmHg ±10 mmHg
2 min ±1 sec
3 min ±1 sec
273 ±3 mmHg
300 ±30 mmHg
10 to 150 mmHg ±3 mmHg
APPA