Purge Sensor Function

The etCO
/Respiratory Mechanics pod accommodates automatic or manual purging
2
of the flow sensor tubing. While the pod automatically purges the breathing circuit
periodically; manual purges are sometimes necessary.
NOTE: Pressure and flow parameters are frozen and waveforms are not drawn during
purging.
Automatic Purging
The following table describes automatic purge characteristics.
Mode
Adult
Adult/Pediatric
Pediatric
Neonatal
1
NOTE: Sensor tubing lines are purged at required interval regardless
of circuit pressure.
Manual Purging
The
and
Paw
Vent
the automatic purge cycles. See "Quick Reference Table: Paw and Vent Setup" on
page 21-8.

Paw and Vent Setup Menus

On the Airway Pressure (
configure the monitor to display respiratory mechanics parameters.
To access respiratory mechanics parameters setup menus:
Click on the parameter box labeled Paw on the main screen to open the
Airway Pressure setup menu or Vent to open the Flow/Volume setup menu.
Each of these setup menus has a second "page" that is displayed by clicking on the
arrow at the bottom of the screen.
more
VF8
D
Ventilator Circuit
Pressure
0 - 40 cmH2O
41 - 60 cmH2O
> 60 cmH2O
Not applicable
Not applicable
menu and
Purge Sensor
) and Flow/Volume (
Paw
/D
XL/K
ELTA
ELTA
P
AW AND
Interval
10 minutes
5 minutes
2 minutes
1
5 minutes
1
3 minutes
function allows manual purging between
) setup menus you can
Vent
APPA
V
S
M
ENT
ETUP
ENUS
21-7