Dräger Infinity Delta Instructions For Use Manual page 552

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

B T
D
ECHNICAL
Relative humidity:
Atmospheric pressure:
Infinity® PiCCO Pod
User Interface
User Controls:
Displays:
Connections:
Degree of protection against ingress
of water:
Physical Attributes
Size (H x W x D):
Weight:
Note: Weight includes two transducer adapter blocks and exclude mounting clamp.
Electrical Specifications
Power source:
Protection against electric shock:
Mode of operation:
Defibrillation protection:
Environmental Requirements
Temperature range:
Relative humidity:
Atmospheric pressure:
B-26
ATA
Operating:
Hemo2, Hemo4, HemoMed: 20 to 90 %, non-condensing
MPod – QuadHemo: 10 to 95%
Storage: 10 to 95 % (with packaging)
Operating:
Hemo2, Hemo4, HemoMed: 525 to 795 mmHg
(70 to 106 kPa)
MPod – QuadHemo: 480 to 795 mmHg (64 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
Fixed keys (C.O. Start, IBP Zero)
four 4-character LCDs.
4 invasive pressures, C.O., 2 temperatures, single cable
connecting pod to monitor
IPX0 (per IEC 60529)
140 x 205 x 60 mm (5.5 x 8.1 x 2.3 in)
0.9 kg (1.9 lb)
Powered directly from the monitor
Type CF
Continuous
Per IEC 601-2-34
Operating:5 to 45 °C (41to 113 °F)
Storage: –20 to 60 °C (–4 to 140 °F)
Operating:10to 95 %, non-condensing
Storage: 10 to 95 % (with packaging)
Operating:485 to 795 mmHg (65 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
D
/D
ELTA
XL/K
ELTA
APPA
VF8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents