Dräger Infinity Delta Instructions For Use Manual page 564

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

B T
D
ECHNICAL
Degree of protection against
electric shock:
Defibrillation protection:
Invasive Blood Pressure (IBP)
Measurement method:
Display resolution:
Measurement range:
Frequency ranges:
Accuracy:
Zero balance range:
Transducer specifications:
Degree of protection against electric
shock:
Defibrillation protection:
B-38
ATA
Type CF
per EN 60601-2-30 (IEC 60601-2-30)
D
/D
ELTA
Resistive strain gauge transducer
1 mmHg
–50 to +400 mmHg
DC to 8 Hz, DC to 16 Hz, and DC to 32 Hz
(user selectable)
±1 mmHg or ±3 % exclusive of transducer
(whichever is greater)
±200 mmHg
Dräger provided transducers with a resistance of 200 to
3000 Ω and an equivalent pressure sensitivity of 5μV/V/
mmHg ±10 %
Type CF
per IEC 60601-2-34
XL/K
ELTA
APPA
VF8